Parameter |
Clinical Meter |
Range |
10 to 299mmol/l |
Sample volume |
20µl or 100µl |
Reproducibility |
±1% for 100µl sample at 100mmol/l, ±1.5% for 20µl sample at 100mmol/l |
Linearity |
<±1mmol/l or ±1% of concentration value (whichever is the greater) |
Power |
230, 50Hz |
Size (w x d x h) |
240 x 215 x 160mm |
Weight |
3.1kg |
Parameter |
Industrial Meter |
Range |
10 to 999mg/l |
Sample volume |
500µl |
Reproducibility |
±3mg/l |
Linearity |
±3mg/l or 1% (whichever is the greater) |
Power |
230, 50Hz |
Size (w x d x h) |
240 x 215 x 160mm |
Weight |
3.1kg |
Part Code |
Description |
504 053 |
Pack of 3 silver anodes |
504 052 |
Pack of 3 silver electrodes (1 cathode, 2 detectors) |
060 028 |
Electrode polish |
025 011 |
Acid buffer solution (500ml) |
025 013 |
Chloride standard, clinical, 100mmol/l (180ml) |
025 014 |
Chloride standard, industrial, 200mg/l (180ml) |
060 029 |
Stirrer bars (pack of 10) |
024 018 |
Glass beaker |
025 012 |
Gelatin (30ml) |
025 195 |
Combined acid buffer (500ml) |